Trade name
Pyrolizer Pyroprobe 5000
Technical description

Pyrolysis is heating large complex molecules in the absence of air to break them down into smaller, more analytically available fragments. Combined with GC/MS, pyrolysis enables quantitative and qualitative analysis of complex samples without extended extraction or derivatization procedures.

In addition, the device offers multi-stage temperature programming. The first step of the analysis temperature below 350 °C allows the extraction of volatile and semi-volatile compounds for analysis before pyrolysis of the remaining polymer. This is invaluable for detecting solvents as well as residual polymer monomers and additives

Conditions for providing infrastructure

In accordance with the Laboratory Regulations available on the website
http://www.orgchem-lab.agh.edu.pl/

Type of accreditation / certificate:
Not applicable
Access type
External
Research capabilities

Analysis of any sample matrix, VOC analysis in air by desorption of sorbent tubes. Ability to work in a wide range of temperatures up to 1400 °C with an accuracy of 1 °C and impulsive or programmable temperature increase.

Last update date
March 8, 2025, 10:29 a.m.
Year of commissioning
2012
Measurement capabilities

VOCs analysis in the air, soil and rock samples, polymer composition analyses, composition of kerogen and asphaltenes analyses.