Differential scanning calorimeter SDT Q600 by TA INSTRUMENTS

Trade name
Differential scanning calorimeter SDT Q600 by TA INSTRUMENTS
Technical description

Analyzer for DSC differential scanning calorimetry and TG thermogravimetry.

Conditions for providing infrastructure

The equipment is made available on the terms resulting from the Regulations for the Use of the Research Infrastructure of ACMiN. (https://acmin.agh.edu.pl/acmin/dokumenty/).

Type of accreditation / certificate:
Not applicable
Access type
External
Research capabilities

The system allows simultaneous TG/DSC analysis. Study of sample weight changes and accompanying thermal effects occurring during dynamic heating/cooling of the sample and/or in isothermal conditions in a static, dynamic, neutral, oxidizing or reducing atmosphere. TG/DSC: 25-1500°C conditions: static, dynamic; atmosphere: inert, oxidizing, reducing. TG/DSC-MS enables qualitative and quantitative analysis of released gases.

 

Last update date
Aug. 30, 2023, 1:26 p.m.
Year of commissioning
2013
Measurement capabilities

WORK PARAMETERS

sample size up to 200 mg

working temperature up to 1500°C

heating rate from 0.1 to 100°C/min

sensitivity 0.1 µg

thermocouple: platinum/platinum-rhodium

DTA sensitivity 0.001°C

calorimetric accuracy ±2%

atmosphere: inert, reducing, oxidizing

Photos
Analyzer for DSC differential scanning calorimetry and TG thermogravimetry.
Analyzer for DSC differential scanning calorimetry and TG thermogravimetry.